Report No 094061R-ITCEP07V06
Page: 94 of 123
10.4. Test Procedure
The EUT is placed on a table that is 0.8 meter height. A ground reference plane is placed on
the table, and uses a 0.1m insulation between the EUT and ground reference plane.
The minimum area of the ground reference plane is 1m*1m, and 0.65mm thick min, and
projected beyond the EUT by at least 0.1m on all sides.
Test on I/O and communication ports:
The EFT interference signal is through a coupling clamp device couples to the signal and
control lines of the EUT with burst noise for 1minute.
Test on power supply ports:
The EUT is connected to the power mains through a coupling device that directly couples the
EFT/B interference signal.
Each of the Line and Neutral conductors is impressed with burst noise for 1 minute.
The length of the signal and power lines between the coupling device and the EUT is 0.5m.
10.5. Deviation from Test Standard
No deviation.