Report No.: 048L022EI
Page: 11 of 35 Version:1.0
3.5. Test Procedure
The maximum spectral power density shall be determined using a spectrum analyzer of adequate
bandwidth for the type of modulation being used in combination with an RF power meter.
Connect an RF power meter to the IF output of the spectrum analyzer and correct its reading using a
known reference source, e.g. a signal generator.
The above procedure shall be repeated for each of the three frequencies identified by the procedure
given in limit (subclause 5.7.2.2.)
Where the spectrum analyzer bandwidth is non-Gaussian, a suitable correction factor shall be
determined and applied.
Where a spectrum analyzer is equipped with a facility to measure power density, this facility may be
used instead of the above procedure.
3.6. Uncertainty
The measurement uncertainty is defined as ± 1.27 dB
3.7. Test Specification
According to ETSI EN 300 328:V1.4.1 (2003-04)
3.8. Test Result
The emission from the EUT was below the specified limits. The worst-case emissions are shown in
section 8. The acceptance criterion was met and the EUT passed the test.